Terms:
·
Gas regulator: pressure-reducing valve that uses opposing
forces on two side of a diaphragm. Regulators are used in
piping systems, on cylinders and inside ventilators to:
o
Maintain a constant pressure because the pressure from the
tanks will decline with use
o
Reduce a high pressure to a lower one. Many regulators
reduce the cylinder's 2000 psig to the 50-60 working
pressure needed for respiratory care equipment
·
Direct action regulators: this regulator operates with a
spring-loaded diaphragm that opposes the pressure of the
gas. The opposition of the force of the spring and the
pressure of the gas keeps the pressure down stream from the
regulator at a constant pressure.
o
When the poppet is at rest, it is closed against the door
that leads to the higher pressure chamber. The lower
pressure chamber starts to empty
o
Once the pressure in the lower chamber drops below the
spring tension, the diaphragm moves the poppet into the
high-pressure chamber. This opens the door. Gas flows from
the higher pressure chamber into the lower pressure chamber
until it equals the spring tension.
o
Then the poppet closes and the flow stops
o
To keep the pressure constant: if the pressure upstream
drops, the poppet opens to allow gas to enter until the
pressure equals the spring tension
o
The diaphragm position and its regulation of the pressure
are determined by the position of the spring.
§
Indirect action regulators: works the same way, the gas
flows differently
§
Single stage regulators: the regulator reduces the pressure
only in one step
§
One determines how many stages a given regulator has by
counting the pressure relief valves.
§
Dual stage regulators: the regulator reduces the pressure
with two single stage regulators in series so that the
pressure goes from highest pressure to a working pressure.
This will have two pressure relief valves
§
Multistage regulators: more than two single stage regulators
in series will have as many stages as there are pressure
relief valves.
§
Adjustable regulators: Some springs are preset at the
factory and others are set by the operator. The regulator is
adjustable because the spring is attached to a handle that
can be turned to tighten or loosen the tension
§
Preset regulators: will not have the adjustment handle. The
tension of each spring will never vary.
§
Pressure relief valves: when the pressure in a regulator
exceeds the appropriate pressure by 50%, the spring disc
opens to vent the chamber to the atmosphere. There is one
pressure relief valve per reducing stage. Over-pressure
tends to happen when there is trash downstream from the
regulator causing decreased flow out of the regulator.
§
Pressure gauges: manometers are devices used to measure
pressure in line or in tanks. They are generally calibrated
in psig or in kPa.

Structure and function of the
pressure gauges called the bourdon gauge
Bourdon gauge:

Bourdon Gauge
The Bourdon gauge works on the same
principle as that of the snakelike, paper party whistle you
get at a New Year party, which straightens when you blow
into it.
Within the Bourdon gauge is a
thin-walled metal tube, somewhat flattened and bent into the
form of a C. Attached to its free end is a lever system that
magnifies any motion of the free end of the tube. On the
fixed end of the gauge is a fitting you thread into a boiler
system. As pressure increases within the boiler, it travels
through the tube. Like the snakelike paper whistle, the
metal tube begins to straighten as the pressure increases
inside of it. As the tube straightens, the pointer moves
around a dial that indicates the pressure in psi.
The Bourdon gauge is a highly
accurate but rather delicate instrument. You can easily
damage it. In addition, it malfunctions if pressure varies
rapidly. This problem was overcome by the development of
another type of gauge, the Schrader. The Schrader gauge is
not as accurate as the Bourdon, but it is sturdy and
suitable for ordinary hydraulic pressure measurements. It is
especially suitable for fluctuating loads.
In the Schrader gauge, liquid
pressure actuates a piston. The pressure moves up a cylinder
against the resistance of a spring, carrying a bar or
indicator with it over a calibrated scale. The operation of
this gauge eliminates the need for cams, gears, levers, and
bearings.
History of the bourdon gauge:
In 1849 the Bourdon tube pressure
gauge was patented in France by Eugene Bourdon. It is still
one of the most widely used instruments for measuring the
pressure of liquids and gases of all kinds, including steam,
water, and air up to pressures of 100,000 pounds per square
inch. Eugene Bourdon founded the Bourdon Sedeme Company to
manufacture his invention.
The American patent rights to Eugene
Bourdon's gauge were bought by Edward Ashcroft in 1852.
Ashcroft played an important role in the widespread adoption
of steam power in this country. He renamed Bourdon's gauge,
the Ashcroft gauge.
Regulators
Regulator with bourdon gauge used as flowmeter

Can the bourdon gauge be used as a flowmeter?
·
Some bourdon gauges have soft copper coils more responsive
to lower pressure changes. If these are attached to a dial
face that is calibrated in liter/minute, the pressure
changes can be converted into a liter flow.
o
For it to work properly to display the flow rate, it must
have a fixed orifice downstream from the bourdon gauge.
§
A fixed orifice would allow more gas through it as the
pressure rises.
§
The gauge actually measures the back-pressure created by the
restricted orifice.
Is the Bourdon gauge an accurate flowmeter?
The bourdon gauge is an accurate flowmeter as long as there
are no restrictions down stream. If the line downstream from
the regulator was to become occluded the pressure would
rise. This rise in pressure would straighten out the copper
coil and an erroneously higher flow rate would be displayed.
Is the pressure accurate in all
conditions?
Bourdon gauges on ventilators are exposed to rapid pressure
changes. At a respiratory rate of 12 bpm the gauge is rising
and falling every 5 seconds. Bourdon gauges will display
less accurate pressures in the face of respiratory rates in
excess of 12 bpm.
Over the last few years, bourdon
gauges on ventilators are being replaced by digital
manometers with a faster response time. "If the digital
manometer has been calibrated, it will always be more
accurate than any bourdon gauge."
·
Does this device work upside down? Yes, there is no problem
with changes in gravity like there are with the other common
type of flow meter
·
Could Bourdon gauges work in the international space lab?
Yes, as long as the cab was pressurized to 760 mmHg [one
atmosphere] & if the space lab were heated to room
temperature.
Explain the structure and the function of the Thorpe Tube.
In this device a variable orifice
and a constant pressure will control the flow rate. The 02
flowmeter must be placed downstream from a regulator that
has reduced the pressure to 50 psig. At a constant pressure,
the gas flow will rise as the orifice opens to allow more
gas to enter. The orifice is controlled by a needle valve.

Thorpe Tube
·
The actual flow rate is measured by the rising of a float.
The gas molecules push up on the float until the tube widens
enough for them to get around it. As the flow rate
increases, the float rises higher. As long as the flow is
stronger than gravity the float will stay suspended.
·
Can this Thorpe tube be used as a flow meter? Yes, if it is
at sea level and room temperature and the appropriate gas is
used.
·
Is the flow rate accurate in all conditions? No.
·
Different gases have different densities and these densities
will cause inaccuracies. Flow meters are gas-specific. An 02
flowmeter cannot be used for compressed air or for C02
delivery. If one uses an 02 flowmeter to deliver Heliox
[02/He mixture] the RCP must make conversions for accuracy.
·
Multiply the 02 liter flow by 1.8 to correct for 80% Heliox
gas & by 1.6 for 70% Heliox mixture
o
The Thorpe tube must be attached to 50 psig, if it is not,
the flow rate will be inaccurate
o
Does this work up side down? While the Thorpe tube may not
be read when it is turned upside down or on it's side,
because the float will move, its flow rate will be constant,
but one must turn the Thorpe tube back to upright to read
the flow rate.
o
The Thorpe tube would be a poor choice for the international
space lab because it requires gravity to work. While the
needle valve actually changes the flow rate, the float would
not be responding properly and the flow meter would not be
functional as a monitoring device.
o
One would do better with a bourdon gauge type of flow meter
or one would use a flow restrictor rather than a flow meter.
The flow restrictor has several adjustable orifices that can
be selected by turning a knob.
Also see
Heliox
Trouble shooting the Thorpe tube
|
Problem is:
|
try this: |
|
The station outlet or the
regulator is leaking.
You hear a hissing sound around
the connection
& feel cold gas around the connection.
The flow may or may not enter the
flow meter based on the degree of the leak
|
1. Have you selected the wrong
connector for the gas?
a. Check the DISS connections of
both flowmeter and station outlet
2. Is the quick-connect getting
worn?
a. Get a new flowmeter
3. Is the station outlet
quick-connect getting too old?
a. Call maintenance to replace the
station outlet. |
|
The flowmeter displays inaccurate
reading:
you set it up & check the flow
with an Erie test flowmeter to see that the flow
is off
|
1. Is the flow meter compensated
for back pressure?
a. Occlude the outlet and see if
the flow meter indicates that the flow is higher
than before. If the flow rises erroneously this
is an uncompensated flowmeter. Change to a
compensated flow meter.
2. Is there an obstruction
downstream such as an entrainment device or a
cool humidifier?
a. Remove the obstruction
b. Check that the humidifier's inlet and outlet
are both patent
c. Change from a low flow to a high flow
humidifier
d. Adjust the incoming flow rate based on the
entrainment device's specifications.
3. If the flowmeter is a Thorpe
tube, is it up right?
a. Turn it upright to read &
adjust the flow rate
4. Is the flowmeter running off
50 psig?
a. Use correct reducing valve/
regulator to get to 50 psig
5. Is the flowmeter at sea level?
a. If you are in a unpressurized
air plane cabin, change to a flow-restrictor
type of flowmeter or bourdon gauge type of
flowmeter
6. Is it at room temperature?
a. Temperatures where the
flowmeter are inaccurate would be intolerable
for other reasons so that unlikely
7. Is it attached to Heliox rather
than 02?
a. Calculate the correct flow rate
by multiplying the set flow rate by 1.8 to get
the accurate flow rate. |
|
the gas is leaking from the tube:
you feel a leak or you hear a hiss
around the flow meter's body |
1. the integrity of the tube is
disrupted
a. replace the flowmeter & send it
out for repair |
|
the flow meter is leaking from the
nipple adaptor |
1. Tighten the nipple adaptor
2. check that the nipple adaptor is not
cross-threaded |
Differentiate between compensated and non-compensated flow
meters
·
Compensated Thorpe tube flow meters are compensated for the
back pressure which in the case of the bourdon gauge will
cause inaccurate readings
·
The compensated Thorpe tube will read the accurate flow rate
as restriction downstream from it rises. As the flow is
restricted, the flowmeter will display lower and lower
flows. If there is complete occlusion, there will be a zero
displayed. If the resistance is removed, the flow will
increase.
For example: you place a cool
humidifier on the nipple of a flowmeter set at 5 lpm.
Because the humidifier's outlet is not patent, it causes a
complete occlusion and the flow rate will drop from 5 lpm to
zero.
As you adjust the flow rate nothing
happens; the float stays at the bottom. Once the humidifier
is removed, the flow rate rises back to 5 lpm.
·
The placement of the float inside the Thorpe tube will
determine whether your flowmeter is compensated or
non-compensated.
o
An uncompensated flowmeter will have a float that is
downstream from the valve. This float is subjected to
atmospheric pressure
§
if there is an obstruction down stream such as attaching an
entrainment device onto the flow meter, the back pressure
builds up and effects the float.
§
It will rise & record an erroneously high flow rate because
the density of the gas under the float has been increased.
o
A compensated flowmeter will have a float that is upstream
from the valve so that the float will be in contact with the
wall pressure rather than the atmospheric pressure.
o
A quick way to remember the compensated flow meter:
|
Wall... ball...valve.... patient |
What are the advantages of having a compensated flow meter?
§
The flow is accurate in the face of an obstruction down
stream. Imagine what would happen if your 02 line became
occluded or pinched off? Your patient could become hypoxic.
o
The compensated flowmeter would show that the flow rate is
zero
o
The uncompensated flowmeter would display an erroneously
high flow rate
§
If the ordered flow rate is too high for your device [such
as a cool humidifier or an entrainment device], as you
increase the flow, the indictor would show that the flow is
not increasing as you turn it…this means there is
restriction down stream
How can I tell if I have a compensated flowmeter?
·
Read the label on the flow meter
·
If there's nothing written, you would attach the flow meter
to the wall or the tank. As soon as the inlet is in contact
with the high pressure, the float [ball] will leap to the
top then fall back to the correct flow.
§
The uncompensated flow meter's float will not leap up
Why can you not use an air flow meter for 02 delivery?
·
Because the DISS connection will not allow you to attach the
air flow meter to the 02 outlet
you must not:
·
Because the air flowmeter is dirty and can be a fire hazard
as gas enters the regulator
·
Because Thorpe tubes are calibrated for their gas and the
flows will not be accurate.
Differentiate between flow restrictors and flow meters
When some persons go home on 02, you
may not trust them to not change the prescribed liter flow.
The flow restrictor can be preset & to get a different flow
rate, the home care RCP needs to change the flow
restrictor's outlet.
Some flow restrictors are operated
off 50 psig for the hospital or for cylinders but other flow
restrictors use 20 psig pressures such as used with LOX
[liquid 02 systems].
Flow restrictors are accurate only
at their rated pressures
What would be an advantage to a flow restrictor?
A patient with chronic hypercapnia
cannot change his 02 to a dangerous level where he is no
longer breathing off his hypoxic drive. With a flow
restrictor, the patient will get only his prescribed liter
flow--- not more--- not less
Because they are so simple to
operate, flow restrictors are getting a lot of use in air
transport and these are the flow meters of choice in the
international space station's hospital.
Five types of floats used in Thorpe tubes
·
Ball float: read in the middle
·
Non-rotating float: read at the top
·
Skirted float: read at the top
·
Plumb bob valve: read at the top
·
H type float: read at the top