Compare humidifiers to the aerosol generators
The use of a cool humidifier is determined by the flow rate of
the device because high velocity gas running down the narrow 02
lines creates a great deal of resistance to flow. Back pressure
builds up, and the pop off valve alarms just as if there was an
occlusion to flow.
-
Nasal cannula below 4 lpm
doesn’t need a humidifier
-
All simple masks need a humidifier because the lowest flow
is 5 lpm
-
Partial and non-rebreather
need humidifiers, but if the flow rises too high to accommodate
a patient’s inspiratory demand, a pressure pop-off valve will
alarm. In this case, the RCP should remove the humidifier
because it is more important to get the Fi02 than the humidity
at this point.
In circumstances that demand the use of the larger aerosol hose,
the flow rates that can move down this much larger lumen are
much higher. These devices that use large bore hoses and heated
humidifiers or aerosol generators [large volume] can handle much
higher flow rates. These devices could be considered high flow
systems; but patients breathing through these devices can
sometimes entrain air so that the Fi02 of many of these devices
is still variable to a point.
Why do we have to put humidifiers or aerosol generators on 02
devices?
As the gas gets colder, it holds less water molecules. The
partial pressure of H20 in the air goes down as the temperature
goes down. This is the principle behind rain following a cold
front. The more water vapor in the warm air in front of the cold
front, the more the water rains out of the air.
Heated air will hold a lot of water. This is the principle
behind why Houston is so unpleasant in the summertime. Gas
coming from a compressed gas system will be cold and dry. At the
patient's mouth it is 21-22 degrees C--- So we add humidity to
minimize the damage cool dry gas causes to the airway mucosa
Air inside the lung was heated to body temperature by the nose
and less by the lower airways and is able to hold much more H20
than it did at room temperature. If there was no water in the
air stream, as it heats up from 22 degrees to 37 degrees, it
picks up water molecules from the lung lining.
Secretions get thicker as water evaporates from the mucosa.
Patient loses body heat into cool dry air. This gets critical
in the small infant.
Heated humidifiers have been used to speed up body heating in
hypothermia. It is controversial just how much heated gas is
needed.
Some feel that the gas should be heated to body temperature [37
degrees C] others say that we should over-heat the gas [40-45
degrees]
The principles behind this fact?
Relative humidity:
it is known exactly how much water gas holds at various
temperatures. The water molecules will evaporate as much as
possible at air temperature until it is saturated at that
temperature but the absolute humidity varies.
RH of 100% at 21 degrees C is much less humidity than RH of 100%
at 37degrees C because the absolute humidity differs
drastically.
Absolute humidity:
the exact mg/liter of water that is held by the air. The person
who breathes 100% RH at 21 degrees will only have 18.4 mg/Liter
of water, while the person who breathes 100% RH at 37 degrees C
[98.4 degrees F] will have 43.9 mg/Liter of water in the air.
Humidity deficient:
the difference between what the gas holds at the nose and the
amount of water the heated gas holds down in the air way is the
deficient. This difference must be made up by evaporation of
water from the body.
At 21 degrees C the absolute humidity is 18.4 mg/liter and the
body at normal body temperature of 37 degrees C needs 43.9
mg/liter. 43.9- 18.4 = 25.5 mg/liter of water will be pulled out
of the mucosa. This is the humidity deficient
How does the humidifier operate?
Gas moves through a device and is exposed to H20 vapors so that
it can pick up as much water vapor as possible. The speed of the
gas flow, and its temperature will both alter the ability of a
gas to saturate with 02 at that particular time.
Humidifiers are rated for the flow rates they can accept, both
for the changes in the ability of a given humidifier to add
water effectively as well as the resistance to flow of the
device. Both efficiency and resistance to flow are limitations
of various types of humidifiers.
The ability of water to enter the gas flow can be improved by
the addition of more surface area for the gas and the water to
interface for evaporation. Other humidifiers will use heaters to
warm the air so that the gas will accept even more humidity.
Cool
Humidifiers
-
Various types of cool humidifiers.
There are four types of cool humidifiers for use with low flow
02 devices
-
The bubble humidifier: the gas flow enters a
tube to pass under the surface of the water. The gas bubbles
pick up water vapor as they move upward toward the outlet.
-
The bubble diffuser humidifier: a diffuser is
added to the bottom of the tube to make the air bubbles that
move through the water even smaller for increased surface area
for water vapor to be added to the gas flow
-
The jet diffuser: this device adds a jet
below the surface of the water that will increase the gas flow
and pull in more water for exchange. It still uses the diffuser
down stream
-
The jet humidifier: this device has a jet port
above the surface of the water. As the gas velocity speeds up,
water is drawn up a capillary tube to be shattered against a
baffle, the water inside this device is suspended inside the
reservoir as a mist that increases the relative humidity of the
gas that passes through the mist.
Devices that operate with cool humidifiers. All
these devices have small, thin 02 lines
Devices that use a heated humidifier or an aerosol
generator. All these devices have corrugated hoses attached to
them called an aerosol hose.
-
Tracheostomy collar
-
Face shield
-
Aerosol mask
-
Brigg’s adaptor [T-tube]
Large volume aerosol generators
The uses of bland aerosols [those that do not contain
medication] include the delivery of sterile water, isotonic,
hypotonic or hypertonic saline to the airways on a continuous
basis with or without artificial airways
Bland aerosols may be given with Fi02 0.21 or more
Bland aerosols may be given continuously or administered for 10
to 20 minutes at scheduled intervals
Indications for large volume aerosols
· Cool mist to the upper airways
-
Via an aerosol mask or a face shield
Post-op patients get cool mist to reduce the irritation that
can result in swelling and edema from extubation. In this
circumstance we use sterile water but sterile saline is less
irritating to those who wheeze
-
Croup, laryngitis or laryngotrachitis ----as long as they
don’t wheeze.
Induce coughing for sputum sample in persons whose secretions
are too scanty to produce easily.
-
Very common for cytology studies to find lung cancers.
Hypotonic or hypertonic saline solutions are used to induce
coughing because they are so irritating
Heated aerosols: Ideally
once the patient’s airway has been bypassed, a heated aerosol
might seem to be preferred to avoid the humidity
deficient, but these devices can over-hydrate patients. More
properly these folks might be better served with a heated high
flow humidifier---but heated humidifiers are more expensive.
Cautions & contraindications to all types of aerosols but
particularly cool ones
-
Any aerosol that is not a bronchodilator has the
potential to cause reflex bronchospasm.
-
Aerosols carry microbes down to the lower airways
-
They are associated with the spread of infections to
caretakers [TB]
-
They are irritating to some people
NOTE
Try to induce a cough without an aerosol first---you would be
surprised how often no one has asked the patient to cough or no
one has taught the patient how to cough into a sputum cup.
SOB---
Check the patient’s history for asthma or COPD before giving the
standard 40%-60% aerosol, or if he presents with SOB, and
wheezing after the addition of the aerosol, you should change 02
deliveries to device with cool humidifier.
Some facts about aerosol generators:
They create particles of water droplets
The 02 devices attached to aerosol generators have large aerosol
hoses that will require draining every 3 hours. More often if
heated aerosol
Fi02 adjustment on pneumatic aerosol generators [the most
common] are forms of entrainment devices-like the entrainment
mask.
As the Fi02 is raised the total flow rate to the patient drops.
As Fi02 drops the total flow rate increases.
The ratio is identical to that of entrainment mask [venturi-mask]
As water builds up in the tubing, there is increased pressure
inside the entrainment device so that the lateral pressure rises
and less air is entrained.
The 02 delivery devices all share the same generator and
entrainment port, but the actual interface [Brigg’s adaptor,
aerosol mask or face shield] has its own entrainment problems so
the delivered Fi02 can vary.
Fi02 on these devices will vary from 0.28 to 1.0 based on the
actual entrainment ports of the aerosol generator
Heated humidifiers
When do we use the heated humidifier?
-
This is the humidification of choice:
-
for the person on mechanical ventilation. We
never use aerosols [heated or cool] on ventilators
-
persons on cool aerosols who begin to cough or
wheeze with the cool mist
-
all persons with fluid retention problems can
develop problems, particularly infants and small children who
can quickly become fluid overloaded.
-
Newborn
infants.
If the baby is small enough to be placed on a warmer or in an
isolette he needs heated humidification---not cool aerosol.
-
persons with artificial airways whose cough is
ineffective, use care with the addition of heated or cool
aerosol because the sudden thinning of secretions can occlude
their airways.
-
These persons frequently require less suctioning if they are
given heated humidifiers rather than heated aerosols.
Heated high flow humidifiers
are capable of delivering a high gas flow at or near body
temperature at a RH of 100%. These are used with the ventilators
and with the same devices that one would use a large volume
aerosol generator with the exception of the tot hut and the 02
tent
-
Heated humidifiers are used with artificial airways
-
Any time that a heater is used there must be an air temperature
probe or a thermometer at the patient's end of the tube.
-
The thermometer should be accurate within +/- 3 degrees
-
The RCP should chart not only the Fi02 and the set flow rate
[on the flow meter] but also the temperature.
-
The temperature should able to get to at least 30 degrees C
[best is 32-34 degrees C ]
-
Should alarm at 40 degrees or more
Other aspects of the
heaters
Servo-controlled
heaters have two thermometers. One at the outlet of the heated
humidifier and one at the patient interface. These exchange data
and the heater is increased or decreased based on their data.
· Getting the sensors wet will cool them off and they may not work
properly
o
Pull them out and dry them off
o
Try to keep the thermometer probe above the gas flow rather than
on the bottom where water can collect
· Servo controlled heaters can be used with heated wire circuits
but these are usually limited to use with newborns who
absolutely must have exact temperature control
· This heated wire will limit rain out of cooled-off water from
the circuit
o
Works well in neonatal circuits and one need only drain the hose
two or three times a shift
o
Very expensive
o
Adult circuits that use heated wires will still have to be
drained every 3 hours. To some this is rarely worth the money
· Some heated humidifiers have a water line for constant feeding
of water into the reservoir
-
Others may have water lines that need to be opened by the RCP
periodically.
-
Increased chance of infection if the RCP has to open the
reservoir to add water
-
The concha
TM
heater has a water jug that holds about 24 hours worth of water,
that constantly drips into the concha
TM
column's wick.
Some thoughts about the hose
All hoses that come from heated or cool aerosols or from heated
humidifiers will have to be drained every 3-4 hours. If heated
aerosol, check every 2 hours
One can put a water trap into the line to collect rainout. Water
traps must be placed in a dependent portion of the hose so that
water drains into the trap. Full water traps get heavy so
support it and drain it frequently. One popular water trap can
be attached to wall suction so that the water is not a source of
infection for the caretakers
Failure to drain hoses will :
-
Put the patient at risk of drowning in the water
-
Carries infection to the patient
-
Will decrease the flow of 02 to the patient
-
Will interfere with the Fi02 of the entrainment
device upstream
How should the RCP chart 02 that is attached to a heated
humidifier or heated aerosol?
Chart every 3 to 4 hours. Drain the hose first then listen to
the patient’s chest [full hoses will mask breath sounds or even
mimic rhonchi]
Chart the breath sounds, the respiratory rate and any
interventions of yours such as adding water, suctioning the
artificial airway or hyperinflation by bagging.
Chart the HR, and Sp02 when available
Check and chart the set flow rate as well as the set Fi02
Check and chart the temperature at the airway
Various types of heated humidifiers:
There are 5 types of heated humidifiers for high flow 02 devices
Heated Passover humidifiers:
gas goes into the device and flows over the top of the heated
water. Water vapor rises and is carried to the patient. The
heater is an immersion heater. Very common with 02 hood for
babies
Heated bubble humidifier:
similar to the bubble humidifier except a heater is added.
Cascade humidifier:
a variation of the bubble diffuser in that the grid breaks up
the bubbles that go thought the heated water. Cascade I & II are
examples
Wick type humidifier:
a variation of the pass-over heater in which the gas enters a
column that is lined by blotter paper soaked in water. There is
more surface area available for the passing gas to pick up
water. The column or the water may be heated.
Vapor phase humidifier:
a hot plate heats up the water from below, then the hydrophobic
filter allows only water vapor to pass into the gas stream
How does the aerosol generator operate? List the different
types and compare each.
Large volume aerosol generators create particles of water with
various means.
All aerosol generators must have sterile water or sterile
saline
Pneumatic jet aerosol generator: most common
Gas flows along a tube with a jet so that entrainment of water
is possible from the capillary tube rising from the water
reservoir
Baffles breaks up the particles into smaller particles, and the
more baffles in the system the smaller the particles
The average particle size is 2-20 microns with a total output of
30-50 mg/ Liter
½ of the particles will be in the 2-4 micron range so that they
will rain out in the conducting airways
Mainstream:
main flow of gas passes through the jet
Side stream:
jet assembly is created then passed into the mainstream
Other facts about the pneumatic jet
Most operate at 8-12 lpm and use entrainment ports to control
Fi02
The same jet stream that entrains water from the capillary tube
frequently will entrain air into the 02 flow
Heaters can be added to the pneumatic jet to increase delivered
fluid
This is the most common means of 02 delivery for aerosol mask,
face tent, Brigg's T adaptors & trach collars
Ultrasonic nebulizer
These devices are rarely used for continuous deliver of gases,
rather they are used to deliver sterile water or saline for
scheduled treatments of 10-20 minutes
Ultrasonic nebulizers put out 1-10 micron sized particles
[average 3 microns] with an output of 60-100mg/liter
The ultrasonic operates on the following principles
Electric energy goes to the piezo-electric transducer
The ceramic transducer vibrates to produce ultrasonic waves that
creates a geyser of aerosol in the center of the couplant.
Couplant water never touches the patient so it can be tap water.
Couplant water must contain electrolytes so it must not be
distilled water.
Immersed in the couplant water is a thin walled solution cup
that carries sterile water or med
The vibrations continue up the water into the sterile solution
above so that it too vibrates
Never put your finger in the water because you will get burned
If the machine fails to create mist, check the couplant water
level. All metal parts should be underwater [electrodes] so that
the electrical impulses can travel
The mist going to the patient is NOT hot, but the water in the couplant will burn you.
You want to see a nice geyser in the center of the couplant,
just above that you should see the mist
Ultrasonic aerosol will stay inside the solution cup unless it
is pushed out by a external fan or by the addition of compressed
air or supplementary 02
Ultrasonic aerosol mist is a very small and very stable
particle
Identify the quick method to tell if a device uses a cool
humidifier or a heated humidifier or an aerosol generator.
Look at the line:
-
Is it a thin 02 line? Then it can only handle a low flow, so
use cool humidifier
-
If it attaches to a large aerosol hose with a 22 mm diameter,
the devices uses higher flow rates and it needs a heated
humidifier or aerosol generator
Look at the outlet.
-
If you see a visible mist no matter how small, it is an
aerosol
-
If you feel the gas but see nothing it is a humidifier that
carries only water vapor
A quick word about room humidifiers
Room humidifiers are used to increase the relative humidity of
the room. Because we put tap water into these devices, they
should never be placed where the mist hits the patient's
face---- rather aim it at the room. Many of these are not true
humidifiers because most of them send out particulate water.
Only the old fashioned steam vaporizer and the new wick room
humidifier actually sends out water vapor.
NOTE:
recently, the EPA has suggested we fill the room humidifier with
distilled water because the minerals found in tap water get
aerosolized and this could be a health hazard.
The EPA also recommends that we clean these devices with scrub
brushes and a solution of 3% hydrogen peroxide every three days,
as well as change out the water Q day.
The EPA suggests we not raise the RH of the room any higher than
50% because that just encourages microbial/ mold/ bacterial
growth. Unfortunately, RH that is comfortable for people is also
more comfortable for bacteria. Asthma experts also caution
asthmatics to avoid room humidifiers for this reason.
The most common room humidifier used in the hospital is the
spinning disk nebulizer. The disk spins and pulls water up
the hollow shaft where the water is slammed against the baffles
by centrifugal force so that the particles are created. The fan
blades create a wind current that moves the mist into the room.
Another type of room humidifier is the ultrasonic mist room
humidifier which sends the mist not down an aerosol tube but
into the room.
The old fashioned steamer/vaporizers are considered a
danger around children because of the chance of steam or hot
water burns.
There is a new room humidifier on the market that uses wick
technology. According to the EPA, the advantage of this
device is that as the RH in the room rises, the device pulls
less water to evaporate on the wick—it is self regulating. It
also transfers no particulate water which means less microbial
contamination and less mineral deposits. This is a true
humidifier like the old steamer/vaporizers.
Fi02's and the flow rates possible with the following
devices used to deliver 02 to the adult and the larger child:
-
Face tents/ face shields
-
T-tubes /Briggs adapters
-
Trach-collars
These devices can run off a heated humidifier, and could
use an entrainment device to mix the air and 02 but they could
also use an 02 blender to get the Fi02 exact.
When the heated humidifier doesn't use an entrainment device to
mix the Fi02, the set flow rate and the delivered flow rate will
be the same. The 02 delivery devices all share the same
generator and entrainment port, but the actual interface
[Brigg’s adaptor, or aerosol mask or face shield] has its own
entrainment problems so the delivered Fi02 can vary.
Fi02 on these devices will vary from 0.28 to 1.0 based on the
actual entrainment ports going into the heated humidifier [or
aerosol generator]
Which of the above devices could be used to deliver an
aerosolized medication if its hose was attached to a small
volume aerosol generator [Nebulizer]?
Medication can be delivered via an aerosol mask, a Briggs's
T-Tube or a Trach-collar. Everything that can be attached to a
large aerosol generator can be attached to a small volume
nebulizer. [SVN]
Indications, hazards and contra-indications, list
the Fi02s and the flow rates possible with the following devices
used to deliver 02 to the infant and the neonate:
All three of these are enclosure 02 devices. These devices
completely cover the patient's head or even his entire body.
These are used in pediatrics because infants will not wear masks
The degree of control the RCP has over any enclosure 02 is
based on the size of the enclosure, and the number of places
where O2 can leak out. Think of a car air conditioner and the
central air conditioner of your house. Which works faster? The
car's air conditioner, because the enclosure is smaller.
If the enclosure is small, there is a lot of control.
If the enclosure is large there is less control.
The Infant Hood:
02 hoods are indicated for delivery of 02 to the tiniest infants
It covers only the head. If the infant is small enough to need a
warmer or an isolette to keep him warm, he needs this because his
02 will be delivered with fine control [+/- 2%] and fine control
over the temperature.
This can be done with a hood because it is such a small
enclosure so that the Fi02 on the blender can be adjusted by
1-2% increments
If there are heated wires on the humidifier, the temperature can
also be controled to .5 degree with ease
Set average flow rates going into a 02 hood at 8-12 lpm.
The tot hut
The baby who is bigger but not big enough to roll over can be
put under a tot hut. This is a soft plastic box that is
assembled with wire corners and placed over the child's head. If
small enough the baby's entire torso gets under the hut
Larger infants can get quite warm under these huts so heated
aerosol may not be needed.
Flow rates need to be at least 12-15 lpm. Failure to keep flow
high can result in the baby getting hot [even to the point of
seeming to have a fever]
Fi02 is adjusted with entrainment device and as Fi02 goes up and
the flow rate goes down some kids might get hot under the slower
flow rates.
Mist tent/ croup tent/02 tent:
-
are indicated for the delivery of bland aerosol and/or
supplementary 02 to the larger infant.
-
If the infant can sit up and stand up he is too big
for the tot hut & is ready for the mist tent.
-
Mist tents are attached to their own high flow rate, cool
aerosol generator
-
Fi02s are not easily adjustable because the enclosure is too
big and too full of leaks for exact Fi02
-
mist tents run off 50 psig
-
if the mist tent is attached to a flow meter set the flow
meter to FLUSH
-
indication is for croup
-
if the kid with croup doesn't need supplemental 02 one can
attach this to compressed air to deliver the needed mist for
croup
Mist tents are inappropriate 02 devices for newborns because [1]
they are always cold and [2] because they are impossible to
adjust the Fi02 to the exactness needed for a newborn
Compare the Fi02 selection that is possible with a 02 blender to
those of an entrainment device.
-
Blenders can deliver 0.21 to 1.0 Fi02 with 1-2% changes at each
step.
-
Entrainment devices will have a few preset values
-
Both devices will have some added entrainment issues due to the
interface
There will be no addition of flow to the blender. What you set
on the flow meter is the actual flow
Which of these devices can be used to deliver the drug Ribavirin?
The SPAG unit can be attached to an 02 hood, a tot hut or can be
put into a croup tent
While the product literature states it can be used with an
aerosol mask, no self-respecting 2 year-old infant will wear an
02 mask without a fight---so put them inside a croup tent.