INTRODUCTION
The Esophageal Obturator Airway (EOA) is used to help
secure an airway in the unconscious unresponsive patient by
blocking the esophagus and channeling air into the pharynx.
Its use during cardiopulmonary resuscitation (CPR) has
greatly reduced the incidence of aspiration as a
complication. The EOA can be inserted without the need to
visualize the airway anatomy.
DISCUSSION
The EOA is a tube approximately 15 inches in length which
is open at the top and has a blind end at the bottom. There
are several side holes located in the upper third of the
tube which allow air to pass from inside the tube into the
pharynx. There is an inflatable cuff located just above the
bottom of the tube. The EOA also has a specially designed
face mask with a hole in the center for the upper end of the
tube and an inflatable cuff to form a seal around the mouth
and nose.
The EOA's blind end is inserted into the esophagus until
the mask is seated on the face. The cuff is then inflated to
seal off the stomach from the upper airway. Air is then
introduced through the tube while a seal is maintained
around the mouth and nose with the mask. The air then exits
through the holes in the tube and from there into the
pharynx.
The EOA therefore provides a more secure airway by
preventing aspiration of stomach contents and channeling air
into the pharynx.
The EOA is designed and intended for short term use only
and should be replaced by endotracheal intubation when
available. It is recommended that endotracheal intubation be
performed with the EOA in place.
The EOA is designed for unconscious unresponsive patients
over 16 years of age, five feet tall and/or 100 lbs. or more
in weight with a need for a secure airway. It is used in
respiratory arrest with or without cardiac arrest.
The EOA is contraindicated in conscious or semiconscious
responsive patients as it will cause retching and vomiting
by activating the gag reflex. It should not be used in
children under 16 years of age or adults under five feet
tall and of less than 100 pounds in weight because of size
factors. It is contraindicated in patients with known
histories of esophageal disease or when caustic poisons have
been ingested as the pressure from the cuff may rupture or
further damage the esophagus. Long term use is discouraged
as the pressure from the cuff will cause necrosis of tissue.
Insertion of the tube should be done within 15 seconds as
to prevent further hypoxia to the patient. If problems
develop during insertion the tube should be removed and the
patient manually ventilated. Then insertion should be
reattempted.
Complications from use of the EOA include endotracheal
intubation in which case the trachea would be occluded and
no air exchange would take place. Auscultation of the lungs
immediately after insertion while ventilating the patient
will allow the EMT to insure correct placement of the tube.
If no lung sounds are present the tube must be removed
immediately and the patient manually ventilated.
Other complications include damage to or rupture of the
esophagus by the EOA lubricating the tip and not using the
EOA in patients with a known weak esophagus will help lessen
incidence of damage as will short term use. Rupture can be
prevented by using only enough air in the cuff to seat the
EOA and careful monitoring of patient's condition so that
the cuff may be deflated when patient begins to vomit.
The cuff must be deflated prior to removal of the EOA.
Removal of the EOA should be done only with adequate suction
available and the patient placed on his or her side to allow
drainage of emesis as the patient will vomit on removal of
the EOA.
Endotracheal intubation can and should be done prior to
removal of the EOA on all patients that require a continued
airway assistance. Dr. Archer 5. Gordon, M.D.,
Anesthesiologist, past chairman of the American Heart
Association, Committee on CPR and Emergency Cardiac Care,
stated "we found that. if anything, having the EOA there
made it easier to intubate. The trachea was the only place
the endotracheal tube could go."
SUCCESSFUL PERFORMANCE CRITERIA
-
Student will correctly perform esophageal obturator
intubation on an adult intubation model. Students must
perform two successful intubations. Criteria for successful
performance in this station is outlined below.
-
Ventilate the intubation model with a bag-mask unit,
demand valve or mouth-to-mouth.
-
Grasp the patient's lower jaw with left thumb and
index finger. Lift the head and flex the head to the chest.
Hold the tongue so the tip of the airway will not push it
into the posterior pharynx,
-
Lubricate the tube generously and insert it with the
curve up. Advance the tube cautiously until it is seated and
the mask makes a tight fit over the mouth and nose.
-
Ventilate via the adapter on the mask to check for
proper placement. If improperly placed remove the tube and
repeat the procedure to this point.
-
If properly positioned inflate the cuff of the tube
until resistance is met. Amount of air needed to cuff the
tube will depend somewhat on the individual. Do not exceed
3035 cc. of air.
-
Attach the bag-valve-mask unit and ventilate, auscultate the chest for bilateral breath sounds. Maintain
seal between mask and face.
-
Ventilate the patient using the esophageal obturator
airway for one minute.
-
Total allowable time to complete each intubation is 30
seconds.
ESOPHAGEAL OBTURATOR AIRWAY
PROTOCOL
Indications
Contraindications
-
Conscious patient
-
Patient with active gag reflex
-
Patient with known history of esophageal disease
-
Patient that has ingested caustic poison
-
Long term use (Greater than 2 hours)
-
Patient under 16 years of age
-
Patients under 5 feet tall and/or under 100 lbs.
weight
-
Laryngectomy
Insertion
-
Assemble EOA Tube inserted through mask tip lubricated
and balloon inflated to check for leaks
-
Hyperventilate patient with 4 full breaths
-
Position at head Do not hyperextend neck Leave in
neutral or flexed position.
-
Thumb lift jaw and hold tongue
-
Insert tube on right side in position of use
-
Advance tube until mask is seated on face. If problems
arise in attempting to insert, remove tube,
manually ventilate patient and then reinsert.
-
Ventilate patient and auscultate the lungs.
-
If EOA is in proper position, inflate balloon until
resistance is met or to a maximum of 30-35 cc. of air.
Precautions
-
Listen for lung sounds as it is possible to intubate
the trachea and no air would be exchanged
-
Do not over inflate balloon may cause esophageal
rupture
-
Have suction ready for immediate use at all times
-
Deflate balloon and suction patient if patient begins
to vomit
Removal Indication
Steps for Removal
-
Turn patient on side have suction ready
-
Deflate cuff
-
Remove tube in one quick smooth motion
-
Suction and oxygenate patient
COMPLICATIONS
|
PROBLEM |
CAUSE |
SOLUTIONS |
|
Regurgitation with the tube in place |
a. Failure to inflate cuff
b. Faulty one-way valve
c. Insufficient inflation
d. Faulty cuff |
a. Suction and inflate the cuff
b. Suction and re-inflate the cuff. Replace the
one-way valve or clamp with padded kellys
c. Suction and inflate more
d. Replace tube and suction, then reinsert |
|
Ruptured esophagus |
a. Over-inflation
b. Improper storage
c. Turning patient's head with tube in place
d. Removing tube with cuff inflated |
a. Titrate inflation according to
patient need
b. store in configuration it comes in
c. Turn patients body
d. always deflate the cuff before removal |
|
Tracheal Intubations |
a. Hyperextending the head and neck
b. Improper storage |
A. Neutral or slightly flexed
position
b. store in configuration it comes in |
|
Insufficient ventilation |
a. Tracheal intubation
b. Improper seal of face mask
c. Improper storage |
a. Proper training - never
hyperextend
b. Proper mask inflation or replacement
c. Proper storage |